Chapter 14 Safety System and Alarms
Alerts, Warnings and Alarms
Your Animas® pump has a progressive Warnings and Alarms safety system. This means if you
do not confirm the Warning or Alarm, it will progress to the sweep alarm with vibrate within
one hour. At the high volume stage, if you do not confirm the Warning or Alarm, the sweep
alarm will begin and will not stop until appropriate action is taken.
Alerts are automatically displayed to remind you of a function that you’ve set or a condition
that exists.
Warnings are triggered for a variety of reasons. They require you to confirm the warning by
pressing OK and/or taking action to address the warning.
Alarms are triggered by several conditions. All require you to address the alarm by taking
appropriate action in order to clear the alarm condition.
- Indicates that this Alert, Warning or Alarm can play a tune as the initial notification.
Refer to your ezManager® User Guide included with the software.
Alert: Active Basal Program Empty
Cause Empty basal program activated.
Effect No basal deliveries.
Message
Displayed once for 3 seconds and each time manually
awakened.
Action None required but can confirm or select Basal Menu.
Beeps/Vib
User selected, one time and each time manually
awakened. No progression.
Alert: Temp Basal Minimum Rate
Cause Negative Temp Basal activated.
Effect Basal delivery will not go below 0.025 U/hr.
Message Displayed once for 3 seconds.
Action None required.
Beeps/Vib User selected, one time. No progression.
Alert: Suspend (Temp Basal/Combo Bolus Canceled)
Cause Pump suspended.
Effect Any active Temp Basal/Combo Bolus canceled.
Message Displayed once for 3 seconds.
Action None required.
Beeps/Vib User selected, one time. No progression.