Tech Specs
Note: When applicable, testing used 23" Comfort™ infusion set and temperature of 23°C +/- 1°C
Number of Basal Segments: 12 per Program
Number of Basal Programs: 4
Basal Delivery Frequency (minimum): every 3 minutes
Temp Basal Range: -90% to +200%, in 10% increments, OFF
Temp Basal Duration: 0.5 hrs to 24 hrs
Extended Bolus Duration: 0.5 hrs to 12 hrs; 0.1 hr
Battery Type: 1.5 Volt Energizer® AA L91 Lithium or 1.5 Volt AA Alkaline
Number of Batteries: 1
Battery Life, Typical use: approximately 5 to 7 weeks for a lithium battery and approximately
2 to 3 weeks for an alkaline battery
Maximum volume infused under single fault condition: Max 2.0 U
Cartridge Capacity up to: 2.0 mL or 200 units
Continuous Operation, Internally Powered Device
Type BF Medical Equipment (Patient isolated, not defibrillator protected)
Watertight Equipment, IPX8 (protected against the effects of submersion, tested at 12 feet/
3.6 meters for 24 hours).
Storage Conditions:
-4ºF (-20ºC) to +140ºF (+60ºC)
10% to 100% relative humidity, including condensing
50 kPa to 106 kPa
Batteries must be removed during storage periods exceeding 2 weeks.
Operating Conditions:
+40ºF (+5ºC) to +104ºF (+40ºC)
Outside these temperatures, the flow accuracy and time to occlusion could be
compromised.
20% to 90% relative humidity, including condensing
Ambient pressure: 70 kPa to 106 kPa